Skip to Content
Calibration Gas Generators

SemeaTech

Disposable sensor calibrators for HF, HCl and formaldehyde, and a benchtop calibration gas generator for ppm-level multi-point calibration.

Disposable sensor calibrators

HF, HCl and CH₂O at ppm Level

SemeaTech calibrators offer a cheap and reliable source of HF, HCl and CH₂O at ppm level.

Ideal for bump testing, the units can also be used for simple gas generation needs. Each calibrator is disposable and good for 6 months or 200 tests. They are much safer and easier to handle than conventional cylinders, and they can be shipped easily. The ppm output of each calibrator can be read from the display on the top plate.

Each calibrator consists of a gas carrier, an equilibrium gas chamber, an on/off valve and an LCD display. At a given temperature the gas concentration in the chamber is constant and shown on the display. Before a calibration or bump test, allow the calibrator 30 minutes or longer to settle to the temperature of the room where the test will be carried out.

  • HF: 15 ppm or 50 ppm
  • CH₂O: 20 ppm
  • HCl: 30 ppm
  • 6 months or 200 tests
  • No cylinders to store or return
  • For the laboratory and the field
SemeaTech HF sensor calibrator CGG-HF-15

Choose Your Calibrator

SemeaTech HF sensor calibrator CGG-HF-15

HF Sensor Calibrator, 15 ppm

SemeaTech · CGG-HF-15

Hydrogen fluoride, 1 to 15 ppm.

SemeaTech HF sensor calibrator CGG-HF-50

HF Sensor Calibrator, 50 ppm

SemeaTech · CGG-HF-50

Hydrogen fluoride, 1 to 50 ppm.

SemeaTech formaldehyde sensor calibrator CGG-CH2O-20

CH₂O Sensor Calibrator, 20 ppm

SemeaTech · CGG-CH2O-20

Formaldehyde, 2 to 20 ppm.

SemeaTech HCl sensor calibrator CGG-HCL-30

HCl Sensor Calibrator, 30 ppm

SemeaTech · CGG-HCL-30

Hydrogen chloride, 1 to 30 ppm.

Calibrator Specifications

HF 15 ppmHF 50 ppmCH₂O 20 ppmHCl 30 ppm
ModelCGG-HF-15CGG-HF-50CGG-CH2O-20CGG-HCL-30
Part number006-0003-001006-0003-002006-0004-000006-0300-000
Concentration range1 – 15 ppm1 – 50 ppm2 – 20 ppm1 – 30 ppm
Accuracy±15%±15%±15%±15%
Operating temperature15 – 30 °C15 – 30 °C10 – 30 °C15 – 30 °C
Pressure range1 ± 0.1 atm1 ± 0.1 atm1 ± 0.1 atm1 ± 0.1 atm
Humidity range15 – 90% RH15 – 90% RH15 – 90% RH15 – 90% RH
Storage temperature10 – 30 °C10 – 30 °C10 – 30 °C10 – 30 °C
Operating life6 months or 200 tests6 months or 200 tests6 months or 200 tests6 months or 200 tests
Power supply3 × AAA or USB-C3 × AAA or USB-C3 × AAA or USB-C3 × AAA or USB-C
HousingPP container, PA66 capPP container, PA66 capPP housing, PA66 coverPP container, PA66 cap
Weight (nominal)220 g220 g220 g300 g

Humidity non-condensing. Operating life is 6 months or 200 tests, whichever occurs first.

New · Benchtop generator

SemeaTech Calibration Gas Generator

An automated, dynamic calibration gas generator based on diffusion and permeation tube technology.

Trace amounts of the target gas released from a tube are diluted with clean zero air to produce ppm-level calibration gas for the multi-point calibration of gas analysers, sensors and gas chromatographs. The tube sits in a temperature-controlled chamber, so the output stays constant and can be verified with certified gas or colorimetric indicator tubes.

  • No concentration loss from cylinder storage
  • One unit replaces several cylinders
  • No high-pressure cylinders on site
  • Traceable via temperature, flow and emission rate
  • PID temperature control, ±0.1 °C
  • Built-in drying and filtration for zero air
SemeaTech calibration gas generator

Target Gases

Diffusion tubes available in October 2026. Replacement tubes are supplied in packs of 5.

GasConcentrationTube part no.
Formaldehyde (CH₂O)1 – 5 ppm411-0500-030/080
Isobutylene (C₄H₈)10 – 100 ppm411-0400-030/080
Hydrogen chloride (HCl)5 – 10 ppm411-0200-030/080
Hydrogen fluoride (HF)5 – 10 ppm411-0300-030/080
Chlorine (Cl₂)5 – 30 ppm411-0800-030/080
Benzene (C₆H₆) · coming soon1 – 10 ppm411-0600-030/080
Water vapour (H₂O) · coming soon50 – 1000 ppm411-0700-030/080

Generator specifications

  • Part number006-0200-000 / 006-0200-001
  • Gas concentration1 to 1000 ppm, depending on the gas
  • Flow range0.1 to 1.0 SLPM, glass tube rotameter
  • Dilution chamber temperature30 to 80 °C, set and displayed to 0.1 °C
  • Working temperature15 to 30 °C
  • Relative humidity≤85% RH, non-condensing
  • Power supply230 VAC 50 Hz (CGG-HOST-220) or 115 VAC 60 Hz (CGG-HOST-110)
  • Max power consumption660 W
  • Dimensions264 × 140 × 192 mm
  • Weight3.56 kg
  • Warranty12 months, excluding accessories
  • Standard accessoriesTube tweezers, Tygon tubing, air drying tube, spare filter cartridges, spare fuses, power cord, user manual, certificate of conformity
  • Optional upgradesPTFE corrosion-resistant dilution chamber, multi-channel expansion module, external zero air generator
What we do

Specification, Not Order Taking

Tell us the detector, the target gas, the alarm level and the environment. We come back with the part that actually works, and the reason.